EasyManua.ls Logo

Inficon 3000 Micro GC - 3000 Micro GC Guide 4; Changing the Carrier Gas Configuration

Inficon 3000 Micro GC
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 - 1
PN 074-558-P1A
3000 Micro GC Guides
3000 Micro GC Guide 4
Changing the Carrier Gas Configuration
Table 4-1 Changing the carrier gas configuration
Step Description
1
The type of carrier gas used for analysis can be changed through the embedded web page.
NOTE: If a carrier gas type is changed without following this procedure, inverted peaks will
appear in the chromatogram.
2 Open a web browser, such as Internet Explorer.
3 In the address bar, type in the current 3000 Micro GC IP Address. Press Enter.
NOTE: At the factory, the instrument IP address was set to 10.1.1.101.
4 The instrument’s embedded webpage will display. Click the Gas Type tab. The current gas
configuration for installed GC modules is shown. Click Make Changes.
5 When prompted for a User name and Password, type gasconfig for both.

Other manuals for Inficon 3000 Micro GC