EasyManua.ls Logo

Inficon BPG402-SE - Setting the Switching Functions

Inficon BPG402-SE
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
tina46e1-b (2017-12) BPG402.om
The threshold values of the switching functions are set locally on the potentiome-
ters of the gauge that are accessible via the openings on one side of the gauge
housing.
Voltmeter
Ohmmeter or continuity checker
Screwdriver, max. ø2.5 mm
The procedure for setting thresholds is identical for all switching functions.
Put the gauge into operation.
Connect the + lead of a voltmeter to the threshold measurement point of the
respective switching function (Pin 3 or Pin 6).
Connect the lead of the voltmeter to a ground contact nearby (e.g. locking
screw of the connector, vacuum flange or housing of the gauge).
The threshold voltages are referenced to ground (housing, vacuum
connection), not to Pin 5 (common power GND 24
V supply).
BPG402-S
BPG402-SL
BPG402-SD
BPG402-SE
BPG402-SP
max. ø2.5
( )
Setpoint Pin 3
max. ø2.5
( )
Setpoint
A
Pin 3
Setpoint
B
Pin 6
Using a screwdriver (max. ø2.5 mm), set the voltage of the selected
switching function to the desired value U
Threshold
.
There is no local visual indication of the status of the switching functions.
However, a functional check of the switching functions (On/Off) can be
made with one of the following methods:
Reading the status via fieldbus interface, for BPG402-SD [1],
for BPG402-SE [3], for BPG402-SP [2].
Measurement of the relay contacts at the sensor cable connector
with an ohmmeter / continuity checker ( 19, 20).
4.13.1 Setting the Switching
Functions
Required tools
Procedure

Table of Contents

Related product manuals