26
tina60e1-a (2011-04)
Mount the gauge so that no vibrations occur. The gauge
may be mounted in any orientation. To keep conden-
sates and particles from getting into the measuring
chamber preferably choose a horizontal to upright posi-
tion and consider using a seal with centering ring and
filter. If adjustment should be possible after the gauge
has been installed, be sure to install it so that the but-
tons can be accessed with a pin.
Remove the protective lid and connect the product to the
vacuum system.
or
Clamp
Seal with centering ring
Seal with centering
ring and filter
Protective lid
Keep the protective lid.