Activating the Factory Setting (Factory Reset)
All user defined parameters (e.g. zero, filter) are restored to their
default values.
Loading of the default parameters is irreversible.
Loading the default parameters:
1 Put the gauge out of operation.
2 Keep the <ZERO> button depressed for at least 5s while
the gauge is being put into operation (Power ON).
5.5
|
Diagnostik-Port (RS232C-Schnittstelle)
The diagnostic port <DIAG> permits to output the pressure read-
ing and all status information and to enter all settings at the
same time (®&[5]).