6
5. Reinstall the rubber sensor cover by pressing it down rmly
around the edges. Be sure the edges of the cover are at against
the surface of the detector.
Figure 2. Installing the Sensor
Using Your INFICON TEK-Mate
W
ARNING
Do not operate this instrument in the presence of
gasoline, natural gas, propane, or in other
combustible atmospheres.
How To Find Leaks
NOTE: A sudden whipping of the leak detector probe or "blowing"
into the sensor tip will affect the air ow over the sensor and
cause the instrument to alarm.
1. Place the tip of the leak-detector probe as close as possible to
the site of the suspected leak. Try to position the probe within 1/4
inch (5 mm) of the possible leak source.
2. Slowly (approximately 1 to 2 inches/second (25 to 50
mm/second)) move the probe past each possible leakage point.