12 tinb07e1-e 2011-07 Vacuum Gauge Controller
3.3.2 Mains connection
The mains connection (Fig. 3-4,  11, Pos. G) is 
designed for a mains cable which contains a European 
appliance connector on the device side.
A mains cable is supplied with the unit. If the plug is not 
compatible with your wall socket, you have to get a suit-
able mains cable:
• Three-conductor cable with protective ground
• Conductor cross-section 3 × 1.5 mm
2
 or larger
Fig. 3-5 Three-conductor cable with protective ground (example)
1 Connect the European appliance connector of the 
mains cord with the mains connection of the unit
2 Connect the plug of the mains cable with the wall 
socket
NOTE:
If the unit is installed in a switching cabinet, the 
mains power can be supplied via a switchable central 
power distributor.
3.3.3 Ground
The ground screw (Fig. 3-4,  11, Pos. I) can be used to 
connect the Vacuum Gauge Controller with the protective 
ground of the pumping station.
1 If required: Connect the protective ground of the 
pumping station with the ground screw. Use a pro-
tective conductor.
3.3.4 SENSOR
The SENSOR connection is used to connect the sensors.
For each channel, there are two connections available 
which are connected in parallel: An 8-pin RJ45 appliance 
socket and a 15-pin D-Sub appliance socket. See 
Fig. 3-4,  11, Pos. A…C.
Pin assignment
Fig. 3-6 SENSOR appliance socket (RJ45)
Fig. 3-7 SENSOR appliance socket (D-Sub, 15-pin)
WARNING
Mains power.
Improperly grounded devices can be ex-
tremely dangerous in the event of a fault.
Use three-wire mains or extension cables 
with protective ground only. Plug the mains 
cable into wall sockets with protective ground 
only.
WARNING
No mains fuse.
The Vacuum Gauge Controller is not 
equipped with a mains fuse.
The wall socket must be protected with a fuse 
(max. 10 A).
1 +24 VDC
2PGND
3 Signal
4 Ident
5 Signal-GND
6 Status
7HV_L
8 HV_EMI
1 EMI-Status
2 Signal
3 Status
4 HV_EMI
5PGND
6n.c.
7 Degas
8 Supply
9n.c.
10 Ident
11 Supply_CDG
12 Signal-GND
13 RXD
14 TXD
15 Chassis
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com