EasyManua.ls Logo

Inficon VGC402 - Pre; Prx; Res; Sav

Inficon VGC402
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40 tinb07e1-e 2011-07 Vacuum Gauge Controller
6.3.22 PRE
Pirani range extension. See Chapter 5.2.12 Pirani range
extension (PrE), 26.
S: PRE[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
6.3.23 PRX
Pressure sensors. Pressure readings of all sensors.
S: PRX<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,±b.bbbbE±bb,c,±d.ddddE±dd,e,±f.ffffEff<CR><LF>
6.3.24 RES
Reset. Reset the serial interface.
Deletes the input buffer. All queued error messages are
sent to the Host. The unit returns to the measurement
mode.
S: RES[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: b,c,d,...<CR><LF>
6.3.25 SAV
Save parameters to EEPROM.
The command SAV,0 resets all parameters to their
default values (factory settings). See Chapter 5.3.4
Default parameters (dEF), 27.
The command SAV,1 saves parameter values which
have been changed via the serial interface in the
EEPROM. These values will be preserved even when the
unit is switched off.
S: SAV,a<CR>[<LF>]
R: <ACK><CR><LF>
Parameter Significance
a Range extension for
Sensor 1
0 = Off (default)
1 = On
b Range extension for
sensor 2 (see above)
c Range extension for
sensor 3 (see above)
Parameter Significance
a Status of channel 1
0 = Measurement data ok
1 = Underrange
2 = Overrange
3 = Sensor error
4 = Sensor switched off
5 = No sensor
6 = Identification error
7 = BPG/BCG/HPG error
±b.bbbbE±bb Pressure reading of sensor 1 in the
current unit of measurement
c Status of channel 2
(see above)
±d.ddddE±dd Pressure reading of sensor 2 in the
current unit of measurement
e Status of channel 3
(see above)
±f.ffffE±ff Pressure reading of sensor 3 in the
current unit of measurement
Parameter Significance
a 1 = Perform reset
b,c,d... Queued error messages
0 = No error
1 = Watchdog control has been trig-
gered
2 = Task(s) not executed
3 = EPROM error
4 = RAM error
5 = EEPROM error
6 = Display error
7 = A/D converter error
8 = UART error
9 = Sensor 1 general error
10 = Sensor 1 ID error
11 = Sensor 2 general error
12 = Sensor 2 ID error
13 = Sensor 3 general error
14 = Sensor 3 ID error
Parameter Significance
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com