24 tinb07e1-e 2011-07 Vacuum Gauge Controller
5.2.2 Gas type (GAS)
Sensors are normally calibrated for a measurement in
nitrogen or in air. The GAS parameter is used to configure
the channel to other gas types.
Tab. 5-4 GAS parameter settings
Cor
If you want to perform pressure measurements in a gas
type without a fixed correction factor, you may multiply
the pressure reading with a variable correction factor. To
this end proceed as follows:
1 Set the GAS parameter to Cor
2 Press the PARA button
The correction factor is displayed
3 Use the arrow buttons the modify the correction
factor
The value of the parameter is changed
The COR indicator is illuminated
You can adjust the correction factor of a sensor in the
range 0.109.99. The setting 1.00 returns the uncor-
rected pressure reading.
NOTE:
The gas type correction is effective only for the fol-
lowing pressures: p < 10
-2
mbar (BPG/HPG/MPG sen-
sor), p < 10 mbar (PCG sensor), p < 1 mbar (BCG
sensor).
5.2.3 Measuring range (FS)
Linear sensors (CDG) require specification of the full-
scale range. You can set this value using the cursor but-
tons. The following values are available:
0.01 mbar
0.01 Torr, 0.02 Torr, 0.05 Torr
0.10 mbar, 0.25 mbar, 0.50 mbar
0.10 Torr, 0.25 Torr, 0.50 Torr
1 mbar, 2 mbar, 5 mbar
1 Torr, 2 Torr, 5 Torr
10 mbar, 20 mbar, 50 mbar
10 Torr, 20 Torr, 50 Torr
100 mbar, 200 mbar, 500 mbar
100 Torr, 200 Torr, 500 Torr
1000 mbar, 1100 mbar
1000 Torr
2 bar, 5 bar, 10 bar, 50 bar
5.2.4 Offset (oFS)
When the offset correction is activated, a previously
specified offset value will be subtracted from each pres-
sure reading. This allows to conveniently measure the
relative pressure with respect to a reference pressure.
The offset correction affects the readings on the display,
the RS232C output, the recorder output, and the switch-
ing functions. The analog outputs, however, are not
affected.
Establishing an offset value and activating the offset
correction
1 Select the oFS parameter
2 Keep the UP button pressed for approximately 2
seconds
The current pressure reading becomes the new
offset value
The offset correction is activated
The OFS indicator is illuminated
Deactivating the offset correction
1 Select the oFS parameter
2 Press the DOWN button
The offset correction is deactivated
The display shows oFF
The OFS indicator is dark
Activating the offset correction
1 Select the oFS parameter
2 Press the UP button
The offset correction is activated
The OFS indicator is illuminated
NOTE:
Always deactivate the offset correction before ad-
justing the offset of a sensor.
Display Significance
n2 Nitrogen or air. No correction of any
kind is required.
Ar Argon. The pressure reading is
determined utilizing a correction fac-
tor for argon. COR is illuminated.
H2 Hydrogen. The pressure reading is
determined utilizing a correction fac-
tor for hydrogen. COR is illumi-
nated.
Cor Other gases. The pressure reading
is determined utilizing a variable
correction factor. COR is illumi-
nated.