13
2. ValiDator Accessories and Application
Before using the ValiDator, it is recommended that the user read this manual
and understand the operation of this system.
List of Equipment & Accessories
User Guide
ValiDator (Calibration Box)
ValiDator Density Data (Laminated Sheets)
Quick Start Instructions
NIST Certificate
List of Optional Accessories
Optional VeriFier Software (USB Flash Drive)
3430 and 3440 Constant Download Cable
InstroTek 3500 Gauge Constant Download Cable
Note: Small scratches should be expected on the ValiDator working surface due
to extensive quality control process utilizing measurements with multiple gauges.
Application Overview
Verification & Calibration Process
There are several steps necessary to accomplish a gauge verification and
calibration. Gauge calibration verification is the most fundamental use of the
ValiDator device. Position a density gauge on top of the ValiDator device, take
a four (4) minute density reading and evaluate the resulting density using the
laminated gauge specific ValiDator Density Data Table provided.
The steps below describe a procedure within the Verification and Calibration
process. Density and Moisture gauge data is required by the VeriFier Software
and should reflect the most recent calibration. If not available, this data may
be obtained from the organization that completed this calibration.