Verification Using VeriFier Software
To utilize the features of the VeriFier Software, you must enter the current gauge
calibration and ValiDator density data into the software. Follow Step I: Entering
Gauge Calibration Data on page 25 and Step II entering ValiDator Densities on
page 26.
Step I: Verification Data Entry
1. Enter the current gauge calibration data into VeriFier software. (Follow
procedure Step I page 25)
2. Enter the ValiDator densities from laminated sheet. (Step II, page 26)
3. Enter counts from ValiDator Evaluation Form taken by the gauge on the
ValiDator.
3.1. Daily Density and Moisture Standard Counts
3.2. Verification Date
3.3. Failure Cutoff, default value is 2pcf (32 Kg/m
3
)
4. Enter Gauge Counts from Evaluation Form
4.1. Enter Moisture Count at Backscatter
4.2. Enter Density Count at all depths
5. Click on Calculate to display the Calculated Density and Error. This density
should be similar to the WD from the ValiDator Evaluation form.
6. To change measurement units, click on Options and select desired units.
7. To print a Verification report, click on Report.
8. To customize your report click on Options, then Customize.