7. Print the Calibration report by Selecting Calibration from the Verifier
screen and click on Report.
Calibration Date
This is the date that the calibration was done.
Gauge Count
Density Counts from each depth taken on
the ValiDator. Depths 0 and 12 are the
average of (2) four-minute readings.
Note: Performing a different calibration method (3 block instead of a 5
block) other than the one used for the current calibration will produce faulty
constants. If an error in the final calibration is noted you may revert back to
the original by pressing CTRL+E.
Daily Standard Count
This is an extended density and moisture
Standard Count from the average of (5) four-
minute standard counts
Calib. Moisture Count
Moisture count taken at backscatter