Supported Nuclear Gauges
It is important when you verify or calibrate your gauge that the correct chart
and values are selected. If there are gauges in your fleet that are not supported
on this list, you may establish guidelines for Verification of these gauges.
(Reference Verification Procedures for Gauges Not Supported page 20)
CPN
MC3 New, MC1 DR New: Gauges manufactured with two GM detectors.
Standard counts are generally between 30000 & 40000 counts.
MC3, MC1 DR: Gauges manufactured before approximately 1996 with one GM
detector. InstroTek currently does not support these gauges. Standard counts
are generally between 10000 & 20000. Please contact InstroTek for more
information on these gauge models or reference Verification Procedures for
Gauges Not Supported page 20.
MC1: InstroTek currently does not support verification of this gauge. Please
contact InstroTek for more information on this gauge model or reference
Verification Procedures for Gauges Not Supported page 20.
MC2: InstroTek currently does not support verification of this gauge. Please
contact InstroTek for more information on this gauge model or refer to manual
section Verification Procedures for Gauges Not Supported page 20.
MC3 and MC1 Elite:
These gauges are supported.
Humboldt
5001 B, 5001C, 5001P, 5001 EZ and 5001 SD are supported.
InstroTek
Model 3500 Xplorer and 3500 Xplorer 2 are supported.
Troxler
3401 & 3411: These gauges are only different in electronic functionality. Both
these gauges are supported by the ValiDator.
3430 & 3440: These gauges are only different in electronic functionality. Both
these gauges are supported by the ValiDator.
3411 New: (Very few gauges were produced, serial numbers greater than 17000)