5. Change data or Select a Different Gauge:
a) From the Gauge pull down menu, select Modify if the gauge data
requires updating or correction.
b) From the Gauge pull down menu, select Pick to review and select a
specific gauge.
CPN MC3, MC1, MC3 Elite and MC1 Elite
If density value for magnesium (low block) provided on the calibration
report is less than 110 pcf (1762 kg/m3), enter the values indicated for all
three blocks on the report for calibration density values. If the values are
110 pcf (1762 kg/m3) or higher, multiply the magnesium (low block)
density by 0.976, magnesium/aluminum (mid block) by 0.954 and
aluminum (high block) density by 0.974 and multiply granite density by 1.
Step II: Entering ValiDator Calibration Densities
(Use the laminated ValiDator Density Data supplied with your ValiDator. These
sheets are manufacturer, model, and ValiDator specific. For large organizations,
a multiple ValiDator function is available. Insure that before a calibration is
attempted that the correct ValiDator is selected.)
1. Select ValiDator from the pull down menu, and Select New. Enter the
ValiDator Serial number, and From the Model pull down, select the
appropriate model from the laminated ValiDator Calibration Table.
2. Enter the Moisture content from the laminated sheet and all the depths and
densities found on your sheet. You may enter Pcf or Kg
3
/m, but not both.
The options menu allows you to change units.
3. To change ValiDator Densities, choose Modify.
4. For selecting another ValiDator, select Pick.