Calibration Procedures
The calibration process utilizes the optional VeriFier Software. Contact InstroTek
to obtain your specific password for operating this program. Note: Only trained,
qualified and experienced users should perform the Calibration and you should
follow the requirements of your nuclear materials license for operating,
maintaining and using your gauge.
The nuclear gauge’s current calibration data is necessary to complete the
calibration process. If not available at your organization, contact the gauge
calibration facility that performed this calibration for a copy.
For loading the Verifier Software see Chapter 3: Software Installation.
Note: If the gauge is not maintained nor has service problems, Calibration is not
going to resolve any service or maintenance problems. Contact your service
center to address any service issues with the gauge.