Setup
3Setup
3.1DealerSetup
WARNING!
Beforethefirstpatientuse,theproviderMUST
disinfecttheproduct.RefertotheDisinfecting
section.
Beforepatientuse,theproviderMUSTcheckthe
hydrostatictestingdate.RefertotheHydrostatic
TestingDatesection.
WARNING!
RiskofInjury
Formedicalapplications,itisveryimportanttoselect
theprescribedlevelofoxygenflow.Uninterrupteduse
ofhighconcentrationsofoxygenoveralongduration,
withoutmonitoringitseffectonoxygencontentof
arterialblood,maybeharmful.
UseONLYasdirectedbyalicensedpractitioner.
DONOTincreaseordecreasetheflowunlessa
changehasbeenprescribedbyyourphysicianor
therapist.
3.2UserSetup
WARNING!
Whenreceivingthecylinderassemblyfromthe
provider,theuser/patientMUSTperformareceiving
inspection.RefertotheReceivingInspectionsection.
3.3ReceivingInspection
Removeproductfrompackageandinspectfordamage.Ifthereisany
damage,DONOTuse.Contactyourdealer/HomeCareequipment
provider.
3.4CylinderPrefillInspection
WARNING!
RiskofInjuryorDamage
Oxygencylinderassembliesshallbeusedforpersonal
useonly.Nottobefilledforresaleoruseby
professionalusers.
FillingofcylinderassembliesisperformedbyInvacare
HomeFillOxygenSystemonly.
ThiscylinderassemblyistobeusedwiththeHomeFill
system.Thiscylindercontains93±3%oxygen
producedbyanoxygenconcentratorusingthe
pressureswingadsorptionprocess.
DONOThandlethecylinderassemblyoruse
contentsuntilyouareprofessionallytrained,including
emergencyprocedures.
Usethiscylinderassemblyinaccordancewiththe
HomeFillusermanual.
Openandclosetheshutoffvalveslowly.Closethe
shutoffvalveaftereachuseandwhenempty.
Securethecylinderassemblyduringstorageanduse.
DONOTdrop.
Nosmokingincylinderassemblyarea.Keepaway
fromheat,flameorspark.
Keepoutofthereachofchildren.
1176098-B-0215