Invacare®ML6,M9andDCylinderwithPostValve
8TechnicalData
8.1TechnicalDescription
TheInvacareOxygenPostValveCylinderAssemblyisanoxygen
storageproductthatcandeliverupto2L/minofcontinuousflow
oxygentothepatientviaanasalcannula.Thecylinderassemblyis
designedtostorecompressedoxygenupto2,100psig(14479kPa).
Thecontinuousflowconserver/regulatorreducesthepressureatthe
cannulaconnectionto50psig(345kPa)andcontainsarotaryflow
dialthatisusedtoadjusttheoxygenflowratetotheprescribed
setting.ThecylinderassemblyisfilledwithUSP93±3%oxygenusing
itsproprietaryfillportthatallowsfillingonlythroughtheuseofan
InvacareHomeFillSystem.Thedurationofoxygenflowbetweenfills
isdependentonthesizeofthecylinderandtheflowsetting.
8.2Specifications
Pressures
InletPressureRange:
300–2100psig
InternalRegulated
Pressure:
50psig
MaxFillPressure:200psig(14479kPa)
BurstPressure:
3025–3360psigat72°F
(20857–23166kPaat22.2°C)
EnvironmentalTemperatureRange
OperatingConditions
Temperature:
–4°Fto140°F(-20°Cto60°C)
StorageConditions
Temperature:
-40°Fto140°F(–40°Cto60°C)
MaximumHumidity:95%RH
AssemblyDimensions
Model:
HF2POST-
6WA
HF2POST-
9WA
HF2POST-
DWA
Weight:4.0lb(1.81kg)5.0lb(2.27kg)6.3lb(2.85kg)
Overall
Length:
11.1in(28.2
cm)
14in(35.5cm)20.8in(52.8
cm)
Diameter:4.4in(11.1cm)4.4in(11.1cm)4.4in(11.1cm)
Cylinders
CylinderSize:
ML6M9
D
Volume(H
2
O):
1L(0.035
ft
3
)
1.7L(0.061
ft
3
)
2.48L(0.088
ft
3
)
OxygenCapacity:
170L(6ft
3
)255L(9ft
3
)422L(14.9
ft
3
)
261176098-B-02