Invacare®ML6,M9andDCylinderwithPostValve
5Maintenance
5.1Maintenance
WARNING!
RiskofInjuryorDamage
Lackofproductmaintenancecanleadtoinjuryandloss
ofproductfunctionordamage.
MaintenanceMUSTbeperformedONLYbyqualified
personnel.
Regularmaintenanceisnecessarytoassureproper
operation.
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausefire.
DONOTuseorfillcylinderassemblyifthecylinder
fillportisdirtyorcontaminated.Contactyour
providerforserviceorrepair.
DONOTcleanwitharomatichydrocarbons,
including,butnotlimitedtobenzeneandtoluene.
DONOTusecleaning/disinfectingsolutionsonthe
fillportfitting.
DONOTimmerseproductinanykindofliquid.
DONOTallowanykindofliquidtogetinside
openingsoftheconserver.
DONOTspraycleaning/disinfectingsolutionson
product.
AllrepairsMUSTbeperformedbyInvacare
CorporationorinanInvacareapprovedrepaircenter.
WARNING!
Riskofinjuryordamage
ThecylinderassemblyMUSTbedisconnectedfrom
theHomeFillandemptiedofgaspriortocleaning
ordisinfecting.RefertotheEmptyingtheCylinder
Assembly.ThefillportcoverMUSTalsobeinstalled.
AllcylindersMUSTbeinspectedbeforeattemptingto
fill;otherwise,injuryordamagemayoccur.Referto
theCylinderPrefillInspection.
Thecannuladoesnotrequiremaintenanceandneedstobereplaced
atthemanufacturer’srecommendedinterval.
Theconserver/regulatorshouldbeinspectedandtestedata
minimumevery5years.ThisinspectionandtestMUSTbeperformed
byInvacareCorporationorbyanInvacareapprovedrepaircenter.
Thealuminumcylindermustundergohydrostatictestingevery5
years.RefertoTheHydrostaticTestingDate.
5.2EmptyingtheCylinderAssembly
DANGER!
RiskofFire
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameor
flammable/explosivesubstances,vaporsor
atmosphere.
DONOTdirectflowofoxygenatanyperson,or
flammablematerialwhenadjustingtheflowdial.
201176098-B-02