Invacare®ML6,M9andDCylinderwithPostValve
5Maintenance
5.1Maintenance
WARNING!
RiskofInjuryorDamage
Lackofproductmaintenancecanleadtoinjuryandloss
ofproductfunctionordamage.
–MaintenanceMUSTbeperformedONLYbyqualified
personnel.
–Regularmaintenanceisnecessarytoassureproper
operation.
DANGER!
RiskofFire
Dirtorcontaminantsonthecylinderassemblyor
conservercancausefire.
–DONOTuseorfillcylinderassemblyifthecylinder
fillportisdirtyorcontaminated.Contactyour
providerforserviceorrepair.
–DONOTcleanwitharomatichydrocarbons,
including,butnotlimitedtobenzeneandtoluene.
–DONOTusecleaning/disinfectingsolutionsonthe
fillportfitting.
–DONOTimmerseproductinanykindofliquid.
–DONOTallowanykindofliquidtogetinside
openingsoftheconserver.
–DONOTspraycleaning/disinfectingsolutionson
product.
–AllrepairsMUSTbeperformedbyInvacare
CorporationorinanInvacareapprovedrepaircenter.
WARNING!
Riskofinjuryordamage
–ThecylinderassemblyMUSTbedisconnectedfrom
theHomeFillandemptiedofgaspriortocleaning
ordisinfecting.RefertotheEmptyingtheCylinder
Assembly.ThefillportcoverMUSTalsobeinstalled.
–AllcylindersMUSTbeinspectedbeforeattemptingto
fill;otherwise,injuryordamagemayoccur.Referto
theCylinderPrefillInspection.
Thecannuladoesnotrequiremaintenanceandneedstobereplaced
atthemanufacturer’srecommendedinterval.
Theconserver/regulatorshouldbeinspectedandtestedata
minimumevery5years.ThisinspectionandtestMUSTbeperformed
byInvacareCorporationorbyanInvacareapprovedrepaircenter.
Thealuminumcylindermustundergohydrostatictestingevery5
years.RefertoTheHydrostaticTestingDate.
5.2EmptyingtheCylinderAssembly
DANGER!
RiskofFire
–NEVERsmokeinanareawhereoxygenisbeing
administered.
–NEVERusenearanytypeofflameor
flammable/explosivesubstances,vaporsor
atmosphere.
–DONOTdirectflowofoxygenatanyperson,or
flammablematerialwhenadjustingtheflowdial.
201176098-B-02