EasyManua.ls Logo

Invacare ML6 - The Hydrostatic Testing Date; External Examination

Invacare ML6
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Invacare®ML6,M9andDCylinderwithPostValve
DANGER!
RiskofInjuryorDeath
Improperfillingorabuseofthiscylinderassembly,or
failuretofollowthesesafetyinstructionsmaycause
seriousbodilyinjuryordeath.Highpressureoxidizing
gasvigorouslyacceleratescombustion.
DONOTalterormodifycylinderassemblyorrelated
components.
DONOTuseoilorgrease.Keepcylinder,shutoff
valve,allequipmentandconnectionsfreeofoiland
greasetoavoidviolentignitionresultinginserious
bodilyinjuryordeath.
3.4.1TheHydrostaticTestingDate
AluminumcylindersMUSTundergotestingeveryfiveyears.
DANGER!
RiskofDeath,InjuryorDamagefromFire
Toavoidfire,death,injuryordamage:
Cylindersthathavebeenrefinishedand/orexposed
toelevatedtemperaturesMUSTbehydrostatically
testedbeforefilling.Cylindertemperatureshouldnot
exceed60°C(140°F).Aluminumcylinderssubjectto
theactionoffireorheatedtemperaturesinexcessof
177°C(350°F)MUSTbewithdrawnfromservice
andcondemnedbytrainedpersonnel.DONOT
usecausticpaintstrippers.Valveandsafetyrelief
devicesMUSTberemovedandreplacedbytrained
andauthorizedpersonnel.DONOTalterorchange
valveassembly.
AllcylindersMUSTbeinspectedbeforeattemptingto
fill;otherwise,injuryordamagemayoccur.
WARNING!
RiskofInjury
Toreducetheriskofinjury:
DONOTfillcylindersthathavenotbeentestedinthe
pasttenyears.Otherwise,seriouspersonalinjurymay
result.Contactyourdealer/HomeCareequipment
providerforreplacement.
3.4.2ExternalExamination
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamage:
DONOTusethecylinderifanyofthefollowing
itemsarefound.Contactyourdealer/HomeCare
equipmentproviderforreplacement.
1.Examinetheoutsideofthecylinderforthefollowingconditions.
Iftheyexist,contactyourproviderforservice.
Dentsordings,arcburns,oilorgrease,anyothersigns
ofdamagethatmightcauseacylindertobeunacceptable
orunsafeforuse
2.Examinethecylinderforevidenceoffireorthermaldamage.
Evidenceincludescharringorblisteringofthepaint,orother
protectivecoatingorheatsensitiveindicator.Iffireorthermal
damageisfound,contactyourdealer/HomeCareequipment
providerforareplacement.
3.Inspectthecylinderassemblyforthefollowingconditions.Ifthey
exist,contactyourproviderforareplacement.
Debris,oilorgrease,noticeablesignsofdamage,signsof
corrosion,signsofexcessiveheatorfiredamage,signsof
pressureleaks
161176098-B-02