Invacare®ML6,M9andDCylinderwithPostValve
1.Priortoeachuse,inspectproductforvisibledamage.Referto
theExternalExaminationsection.DONOTuseifanydamageis
found.
Ifforanyreason,anylabelbecomesillegibleorlost,
contactyourdealer/HomeCareequipmentprovider.
2.InspectthecylindergaugeFtoensurethepressureisatleast
1500psig(10342kPa).Ifbelow1500psig(10342kPa),thenrefill
thecylinderassemblyusingtheHomeFillsystem.
WARNING!
–Whencylindergaugeneedleentersredzone,itis
recommendedtorefillthecylinderassemblyon
theInvacareHomeFillCompressor.
3.PositionthecylinderassemblyAsotheoutletconnectionis
pointingawayfromtheuserandanyotherperson(s).
WARNING!
RiskofInjury
Toreducetheriskofinjury:
–DONOTdirectflowofoxygenatanyperson,or
flammablematerialwhenadjustingtheflowdial.
4.TurntheflowdialGtosetting“0”.
5.Removetheoutletconnectioncap,ifpresent.
6.Attachastandardorhighflowsinglelumenadultoxygennasal
cannulawithamaximumlengthof2.1m(7feet)totheproduct’s
outletconnectionaccordingtothecannulamanufacturer’s
instructions.
DONOTusepediatriccannulaswiththisproduct.
DANGER!
RiskofInjuryorDeath
Thecannulacancausetripping,falling,orother
injuryifimproperlypositionedandsecured.To
avoidinjuryordeath:
–ThecannulaMUSTberoutedandsecured
properly.
–DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhilewearing
thecannula.
–Avoidpositioningcannulasacrossareasofhigh
foottraffic(i.e.aisles,doorways,hallways,etc.)
7.Placethecannulaoveryourearsandpositiontheprongsinyour
noseasinstructedbyyourHealthCareproviderorcannula
manufacturer.