EasyManua.ls Logo

Invacare pegasus pro - Page 58

Invacare pegasus pro
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
InvacarPegasusSeries
FlashcodeFault
Consequenceforthe
scooterComments
5
BrakefailureStopsdrivingEnsurethatthedisengagingleverisintheengaged
position.
Thereisadefectinthebrakingcoilorinthecabling.
Checkthemagneticbrakeandcablingforopen
orshort-circuitedcircuitry.ContactyourInvacare
provider .
6
Noneutralpositionwhen
switchingscooteron.
StopsdrivingDriveleverisnotinneutralwhenthekeyswitchwas
turned.Putthedriveleverinneutral,turnthepower
offandthenturnonagain.
Itmaybenecessarytoreplacethedrivelever .Contact
yourInvacareprovider .
7
Faultinspeed
potentiometer
StopsdrivingThedrivelevercontrolscouldbefaultyor
incorrectlyconnected.Checkthecablingforopenor
short-circuitedcircuitry.
Potentiometerisnotcorrectlyadjustedandmustbe
replaced.ContactyourInvacareprovider.
8
MotorvoltageerrorStopsdrivingThemotororitscablingisdefective.Checkthecabling
foropenorshort-circuitedcircuitry.
9
Miscellaneousinternal
fault
StopsdrivingContactyourInvacareprovider .
10
Push/freewheelmode
error
Stopsmoving
Thescooterhasexceededthepermissiblemaximum
speedduringpushingorfreewheeling.Switchthe
electronicssystemoffandonagain.
581638629-A

Table of Contents

Related product manuals