Invacare®PegasusSeries
FlashcodeFault
Consequenceforthe
scooterComments
5
BrakefailureStopsdriving•Ensurethatthedisengagingleverisintheengaged
position.
•Thereisadefectinthebrakingcoilorinthecabling.
Checkthemagneticbrakeandcablingforopen
orshort-circuitedcircuitry.ContactyourInvacare
provider .
6
Noneutralpositionwhen
switchingscooteron.
Stopsdriving•Driveleverisnotinneutralwhenthekeyswitchwas
turned.Putthedriveleverinneutral,turnthepower
offandthenturnonagain.
•Itmaybenecessarytoreplacethedrivelever .Contact
yourInvacareprovider .
7
Faultinspeed
potentiometer
Stopsdriving•Thedrivelevercontrolscouldbefaultyor
incorrectlyconnected.Checkthecablingforopenor
short-circuitedcircuitry.
•Potentiometerisnotcorrectlyadjustedandmustbe
replaced.ContactyourInvacareprovider.
8
MotorvoltageerrorStopsdriving•Themotororitscablingisdefective.Checkthecabling
foropenorshort-circuitedcircuitry.
9
Miscellaneousinternal
fault
Stopsdriving•ContactyourInvacareprovider .
10
Push/freewheelmode
error
Stopsmoving
•Thescooterhasexceededthepermissiblemaximum
speedduringpushingorfreewheeling.Switchthe
electronicssystemoffandonagain.
581638629-A