EasyManua.ls Logo

Invacare Perfecto2 Series - Humidifier Cleaning

Invacare Perfecto2 Series
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
7.3HumidifierCleaningandThermic
Disinfection
Cleananddisinfecttheoxygenhumidifierdailytoreduce
limestonedepositsandeliminatepossiblebacterial
contamination.Followtheinstructionsprovidedbythe
manufacturer.Ifnoneareprovided,followthesesteps:
1.Washhumidifierinsoapywaterandrinsewithasolutionoften
partswaterandonepartvinegar.
2.Rinsethoroughlywithhotwater.
3.Airdrythoroughly.
Tolimitbacterialgrowth,airdrythehumidifierthoroughly
aftercleaningwhennotinuse.RefertoSetUpforuse.
7.4CleaningandDisinfectionBetweenPatients
WARNING!
RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto
concentrator:
Onlyqualifiedpersonnelshouldperformcleaning
anddisinfectionoftheoxygenconcentratorand
accessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossiblepathogen
exchangebetweenpatientsduetocontaminationof
componentsoraccessories.PreventiveMaintenanceshould
alsobeperformedatthistimeifnecessary.
1.Disposeofandreplaceallpatientsideaccessoriesnotsuitable
formultiplepatientuse,includingbutnotlimitedto:
NasalCannulaandTubing
Mask
Humidifier
2.Performmaintenanceproceduresdescribedinthismanualand
itemsonPreventiveMaintenanceChecklist
3.Checkconcentratorforpossibleexternaldamageorsignsthatit
mayrequireserviceorrepair.
4.Ensureconcentratorfunctionsproperlyandallalarmsarein
workingorder.
5.Beforerepackaginganddistributiontonewpatient,ensure
packagingcontentscontaintheconcentrator,powercord,air
inletscoop,assemblyinstructions,humidifier,cannula,labels
andusermanual.