Invacare®Perfecto™Series
10Technicaldata
10.1TypicalProductParameters
ElectricalRequirements:
120VAC+10,–15%(132VAC/102VAC),60Hz
RatedCurrentInput:3A
SoundLevel:43±2dBAAverageforPerfecto
2
ModelsandPerfecto
2
VModel
40±2dBAAverageforPerfecto
2
WModels
Altitude:
AllPerfecto
2
Models:Upto8,000ft(2438m)abovesealevel
withoutdegradationofconcentrationlevels.
AllPerfecto
2
ModelswhilefillingaHomeFillSystem:Upto
6,000ft(1828m)abovesealevelwithoutdegradationofconcentration
levels.
AtmosphericPressureRange:101.3kPa75.0kPa
OxygenOutputConcentrationLevels:87%to95.6%at0.5to5L/min
Concentrationlevelsachievedafterinitialwarm-upperiod
(approximately30minutes)
MaximumOutletPressure:5psi±0.5psi(34.5kPa±3.45kPa)
LowFlowAlarm:0.5to5L/min(maximum).Forflowrateslessthan1L/min,we
recommendtheuseoftheInvacarePediatricFlowmeterAccessory
(IRCPF16).
PotentialObstructionAlert:0L/minto0.5L/min
Theconcentratordetectsaconditionthatmayindicateapotential
obstructionoftheoutputoxygen.Rapidaudiblebeepingalert(thisalert
isdeactivatedwhenaccessoriesareconnected).Maybeassociated
withflowsettingof0.5L/minorless.
301148074-F-06