Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:
PotentialObstructionAlert
Rapid
Beep...Beep...
Beep...Beep
1.Possibleinternalobstructioninthe
oxygenpath.Kinkedorblocked
tubing,cannulaorhumidifier.
2.Flowmetersetat0.5L/minorless.
1.Inspectforkinksorblockages.Correct,cleanorreplace
item.Oncecorrected,turnpowerOfffor60secondsand
thenturnpowerbackON.
2.Resetflowmetertoprescribedflowrate.Ifprescribed
flowrateisbelow0.5L/min,usethePediatricFlowmeter
IRCPF16(OnlywithPerfecto25).
Theuseofthepediatricflowmeterwilldeactivatethe
PotentialObstructionAlert.
Alarm:
SystemFailure:
YELLOWorREDlightIlluminated.
Alarm:ContinuousOnREDlight
*Onlyappliesto
IRC5PO2
1.Lowoxygenpurity.*
2.Kinkedorblockedtubing,cannula
orhumidifier.*
3.Flowmetersetat0.5L/minorless*.
4.Unitoverheatingduetoblocked
airintake.
5.Insufficientpoweratoutlet.
6.Internalrepairsrequired.
1.CleanorReplacefilters.
2.Inspectforkinksorblockages.Correct,cleanorreplace
item.Oncecorrected,turnpowerOfffor60secondsand
thenturnpowerbackOn.
3.Checkflowmeterissetto1.0L/minormore.Referto
TypicalProductParameters.
4.Dooneorbothofthefollowing:
a.Removeandcleancabinetfilters.
b.Moveconcentratoratleast12in(30,5cm)fromwalls,
draperies,andfurniture.
5.DONOTuseextensioncords.Movetoanotherelectrical
outletorcircuit.
6.CallServiceProvider.
GREENlightwithYELLOWLight
Flashing.
1.Internalrepairsrequired.1.CallServiceProvider.
1148074-F-0629