Invacare®Perfecto₂™Series
DANGER!
RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientsonflow
ratesgreaterthan5l/mmaybeatanincreasedriskfor
seriousinjuryordeathintheeventoffailure.
–ALWAYSdiscussthisincreasedriskwithyourhealth
careproviderBEFOREusingthisproductifyouare
prescribedaflowrategreaterthan5l/m.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand
productparametershasnotbeentestedandmayleadto
productdamage,lossofproductfunction,orpersonal
injury.
–DONOTusethisproductinanywayotherthan
describedintheproductparametersandintendeduse
sectionsofthismanual.
DANGER!
RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen
concentratorinthemarkettoday,thisoxygen
concentratorcanfailtoproduceoxygenduetopower
failureordevicemalfunction.
–ALWAYShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient
toswitchtotheirbackupsourceofoxygen.Referto
Troubleshootingformoredetail.
1.3Description
TheInvacarePerfecto
2
concentratorisusedbypatientswith
respiratorydisorderswhorequiresupplementaloxygen.Thedevice
isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87%
to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse
ofanasalcannula.
TheInvacarePerfecto
2
concentratorusesamolecularsieveand
pressureswingadsorptionmethodologytoproducetheoxygen
gasoutput.Ambientairentersthedevice,isfilteredandthen
compressed.Thiscompressedairisthendirectedtowardoneof
twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed
andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe
patient.
TheInvacarePerfecto
2
concentratoriscapableofoperationbythe
patientinahomeenvironmentorinaninstitutionalenvironment.
Deviceoperatesatanominal120VAC/60Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified
technicalpersonnelonly.
1.4Contraindications
InvacareisnotawareofanycontraindicationsfortheInvacare
Perfecto
2
ConcentratorSeries.
61148074-F-06