Section 4: Device Description
27 of 96
Pulse Mode
Repeat mode in Normal or Fast repetition rate or in Off as a
single pulse mode.
Counter Reset
The Counter can be reset.
Pulse Counter
Shows the number of pulses delivered from the beginning of
the treatment.
System Mode
The interfaces have three treatment modes. Standby, Ready,
and Active.
Standby mode allows the user to set treatment parameters.
Activation of energy is not allowed in Standby mode.
In Ready mode, the system is waiting for a signal from the
Trigger Switch to activate the energy. Any attempt to change
the treatment settings switches the system to Standby
mode.
When the signal from the trigger is indicated, the System
switches to Active mode. Any attempt to change the
treatment settings switches the System to Standby mode.
Main Menu
Return to the Main Menu to select another Applicator and
change the Applicator if needed.
Figure -9: Fractora Treatment Screen
The Fractora Treatment Screen contains:
Selection Frame
The frame selects parameters that can be changed by
functional keys.
Tip Type
Allows selecting between different tip types of Fractora.
Energy
Delivered energy is changed from level 5 to 62 energy levels
and the System starts up at the minimal energy setting.