InMode Without Vasculaze Operator Manual
Section 4: Device Description
29 of 96
The Forma/Plus Treatment Screen contains:
Selection Frame
The frame selects parameters that can be changed by
functional keys.
Applicator
Allows selecting between Forma and Plus Handpieces.
Energy
Delivered energy is changed from level 20 to 62 energy
levels and the System starts up at the minimal energy
setting.
Cut-Off
Cut-off temperature settings are changed in the range of 35-
43ºC. When measured skin surface temperature reaches the
pre-set limit the RF energy is cut off.
Pulse Mode
Allows selecting between single RF pulses of 30sec in the
Single Mode and continuous RF in Repeat Mode, as long as
the Footswitch is pressed.
Counter Reset
The Counter can be reset.
Temperature
Measure
This indicator shows the skin surface temperature, as
measured by an integral temperature sensor.
Treatment Time
This function shows the total time that RF is activated.
Main Menu
Return to the Main Menu to select another Applicator and
change the Applicator if needed.
System Mode
The System has three treatment modes, Standby, Ready,
and Active.
Standby mode allows the user to set treatment parameters.
Activation of energy is not allowed in Standby mode.
In Ready mode, the system is waiting for a signal from the
Footswitch to activate the energy. Any attempt to change
the treatment settings switches the system to Standby
mode.
When the signal from the Footswitch is indicated, the
system switches to Active mode. Any attempt to change the
treatment settings switches the system to Standby mode.