InMode Without Vasculaze Operator Manual
Section 12: System Maintenance
Failure to replace your filters can lead to serious malfunctions of the
laser/IPL Handpieces!
Filling Water for Diolaze, DiolazeXL and Lumecca
InMode System, Diolaze, DiolazeXL and Lumecca handpieces are shipped without
water. Distilled water must be used and failure to do so will void the service
warranty.
During water filling, air bubbles can form inside the hoses which may cause water
flow issues. This instruction will help prevent this from occurring.
Connect one of the Handpieces.
Open the back cover and take out the Bottle.
Unscrew the water Bottle Cap (Figure -1).
Slowly pour distilled water into the Bottle while the Water Filter and Tube are
inside the Bottle to prevent over pouring (Figure -2).
Figure 21-1: Water Bottle Cap
Figure 21-2: Distilled Water
Filling
Figure 21-3: Water Bottle in
Bracket
Fill the Bottle up to the maximum level.
Close the bottle cap and insert the Bottle at the back of the System
(Figure -3).
Connect any Handpiece to the System.
Turn the System On.
Select “Utilities” on the Main Menu (Figure -4).
Select “Water Drain/Fill” from the Utilities Screen (Figure -5).