InMode Without Vasculaze Operator Manual
Section 12: System Maintenance
76 of 96
 Pressing the Up arrow will activate the Pump and count down from 600
seconds (10 minutes) to 0 seconds (Figure -6).
Figure 21-4: Menu Screen
Figure 21-5: Utilities Screen
Figure -6: Time Counter Water Pump Current Value
 After 300 seconds (while the Pump is working), unscrew the plastic Cap on the
side of the Capsule Filter counterclockwise to release air. Afterwards, screw
the Cap clockwise to close it (Figure -7). Ensure that no water is leaking.
Figure -7: Capsule Filter with Plastic Cap
 Pressing the Down arrow will stop the Water Pump.
 Turn the System Off.
 Repeat steps 2 to 5 since the System and Handpiece will take some of the
water.