InMode Without Vasculaze Operator Manual
Section 12: System Maintenance
Observe the reservoir; when no more water is pouring out, turn off the
System.
Draining the Laser/IPL Handpieces with the Water Draining
Tool
When only applicators are shipped or stored, follow applicators draining instructions
below. It is possible to connect them to the system and perform the draining
procedure, as above. It is also possible to use the Invasix Water-draining tool (Figure
-10).
Figure -10: Water Draining Tool
The Diolaze, DiolazeXL and Lumecca Handpieces may be damaged in cold (freezing)
temperatures that occur during transportation or the local storage environment. In
cold climates, especially during the winter, systems and handpiece may be damaged
during transportation or storage when exposed to freezing temperature.
To Drain the Handpiece
1. Hold the Water Draining Tool's Connector and insert the Handpiece Connector
with the label at a 90° counter-clockwise angle (as shown ii Figure -11), and
twist it 90° to the right, so that the label is straight.
Figure -11: Connecting the Handpiece to the Water Draining Tool