EasyManuals Logo

Ion Technologies IT- DS05-S User Manual

Ion Technologies IT- DS05-S
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
IONTECH
®
IT-DS EDI Modules
Troubleshooting Flow Chart - Low EDI Product Water Quality
Page 1 of 3
START HERE
EDI Check IP-Pro: Adjust DC amperage
feed TDS Yes is desired quality Yes Up for higher TDS
changed? possible?
No
Reduce TDS
(clean or replace
No RO membranes)
or lower EDI prod
flow
EDI Check IP-Pro: Adjust DC amperage:
feed CO
2
Yes is desired quality Yes up for higher CO
2
changed? possible? down for lower CO
2
No
Increase pH ahead
of RO to remove CO
2
No (contact Tech Support)
or lower EDI prod
flow
EDI
Feed SiO
2
Clean or replace f
RO membranes or
too high? reduce RO recovery
No
Go to next flowchart
37 IT-DS MAN Rev 0

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ion Technologies IT- DS05-S and is the answer not in the manual?

Ion Technologies IT- DS05-S Specifications

General IconGeneral
BrandIon Technologies
ModelIT- DS05-S
CategoryControl Unit
LanguageEnglish