Function and Features
Copyright ©ITECH Electronics Co., Ltd. 24
firstly. When the external factors are excluded, please press [On/Off]
button.Then the unit could have a output voltage again.If in remote control
mode,you should clear the OVP state,then could you open the output by OUTP
ON command.
Set the OVP voltage value as follows:
1. Press the composite key [Shift] + [V-set] (Setup) to enter the setup Menu.
2. Select “ Source” in the menu and press [Enter] for confirmation.
The power supply can set its OVP.
3. Select OVP with Left/Right key and press [Enter] for confirmation.
4. Select On to enable OVP function and press [Enter].
5. Set OVP voltage value with numeric key and press [Enter] for confirmation.
6. Set OVP delay time with numeric key and press [Enter] for confirmation.
Press [Esc] to exit menu setting.
OCP
User can enable the over current protection function and set the protection
value in setup menu, Over Current Protection will be triggered when the current
in circuit exceeds the protection value.
Once the power supply is over voltage protected,will the output be shut down at
once,and “ Prot” indicator lamp will be lit, and prompt “ Over Current” will
be displayed on VFD screen. At same time, the beeper will be on.
Set the OCP current value as follows:
1. Press the composite key [Shift] + [V-set] (Setup) to enter the setup Menu.
2. Select “ Source” in the menu and press [Enter] for confirmation.
3. Select OCP with Left/Right key and press [Enter] for confirmation.
4. Select On to enable OCP function and press [Enter].
5. Set OCP current value with numeric key and press [Enter] for confirmation.
6. Set OCP delay time with numeric key and press [Enter] for confirmation.
Press [Esc] to exit menu setting.
OPP
OPP is a protection measure taken when the actual power exceeds the rated
power of the power supply. Under OPP, the power supply output will be
switched off and VFD indicator “ Prot” will be lighted on. In addition, the VFD
display screen will display “ Over Power” .
This series of power supply can set OPP.
Set the OPP current value as follows:
1. Press the composite key [Shift] + [V-set] (Setup) to enter the setup Menu.
2. Select “ Source” in the menu and press [Enter] for confirmation.
3. Select OPP with Left/Right key and press [Enter] for confirmation.
4. Select On to enable OPP function and press [Enter].
5. Set OPP power value with numeric key and press [Enter] for confirmation.
6. Set OPP delay time with numeric key and press [Enter] for confirmation.
Press [Esc] to exit menu setting.