EasyManua.ls Logo

ITech IT8514C+

ITech IT8514C+
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Input Settings
Copyright © ITECH Electronic Co., Ltd. 33
CURRent: Constant Current Mode
VOLTage: Constant Voltage Mode
POWer: Constant Power Mode
RESistance: Constant Resistance Mode
DYNamic: Dynamic Operation Mode
LED: LED Mode
IMPedance: Constant Impedance Mode
*RST Value
CURRent
Examples
MODE RES
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
<NRf+>
This command is used to set the current setting in CC mode.
Command Syntax
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Parameters
MIN~MAX|MINimum|MAXimum
Unit
A
*RST Value
MINimum
Examples
CURR 5
Query Syntax
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Returned Parameters
<NR2>
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
This command is used to query the current setting in CC mode.
Command Syntax
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?

Table of Contents

Other manuals for ITech IT8514C+

Related product manuals