OCP Testing Commands
Copyright © ITECH Electronic Co., Ltd. 51
Unit
S
Examples
OCP:DWEL 0.01 or OCP:DWEL 10ms
Query Syntax
OCP:DWELl?
Returned Parameters
<NR2>
OCP:DWEL?
This command is used to query the single-step dwell time of OCP test.
Command Syntax
OCP:DWELl?
Parameters
None
Unit
S
OCP:VTRig <NRf+>
This command is used to set the OCP trigger level.
Command Syntax
OCP:VTRig <NRf+>
Parameters
MIN~MAX
Unit
V
Examples
OCP:VTR 11.8
Query Syntax
OCP:VTRig?
Returned Parameters
<NR2>