EasyManua.ls Logo

ITech IT8600 Series

ITech IT8600 Series
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Commands
Copyright © Itech Electronic Co., Ltd. 58
CALibrate:CURRent:LIMit:POINt
This command is used to calibrate the current limit point in CV mode.
Command syntax:
CALibrate:CURRent:LIMit:POINt <point>
Parameter:
P1|P2
Unit:
None
CALibrate:CURRent:LIMit
This command is used to calibrate the current limit value in CV mode.
Command syntax:
CALibrate:CURRent:LIMit[:LEVel] <NRf>
Parameter:
Current< external reading >
Unit:
A

Table of Contents

Other manuals for ITech IT8600 Series

Related product manuals