What to do if my Jackson Laboratory Equipment keeps tripping the service breaker?
- JjoshuaandersonAug 22, 2025
If your Jackson Laboratory Equipment keeps tripping the service breaker, there are several potential causes. First, check for loose wires or burned connections, as the power supply might be shorted to the ground. Next, inspect the door switch or detergent safety switch for loose or wet connections; bypass the switch to see if it's the problem and replace it if needed. Also, a jammed pump impeller can cause this issue, so clear it. If the wash pump motor is faulty, check its voltage and amp load; replace the motor if the amp load exceeds 12 amps. Lastly, ensure the circuit breaker isn't too small by comparing it to the data plate specifications and replacing it with a properly sized breaker if necessary.