What to do if the prime switch does not activate sanitizer pump in Jackson Conserver AXL?
- OOscar WeberSep 12, 2025
If the prime switch on your Jackson Laboratory Equipment isn't activating the sanitizer pump, it could be due to a faulty prime switch. With the prime switch in the prime position, check for voltage between the GREY and WHITE/RED wires to the switch. If 120V is present, replace the switch. Alternatively, a faulty delime switch could be the cause. With the delime switch in the DELIME position, check for voltage between the WHITE/BLACK and WHITE/RED wires to the delime switch. If 120V is present, replace the delime switch.