Vetek Weighing AB
9. Troubleshooting and Error Message
Error Status
Possible Causes Basic Inspection & Treatment
E02 no.samp
PCS no sampling
Press down the key N/G to begin
sampling.
E04 EE.ERR
EEPROM access error
Re-solder EEPROM or
contact
Service .
E05 out.ran
Overload condition
Remove weight that is greater than
the scale capacity from the pan.
E07 E.Lo>Hi
The weight-
limit value is less than the
lower limit value.
Reset the upper limit value
E08 CK.UNIT
The unit set for weight-
checking is different from the
current unit.
Disable weight-checking function
or reset HI/LO value.
EU.nit
improper accumulation unit
Return to the previous
accumulation unit or accumulate
again.