1. Keep pressing until you see the display as shown.
Arrow “ ” pointing at HI means that the first 16 characters can be edited.
2. Keep pressing and releasing . The display will show , , , in order. Choose for
example.
3. Press or do nothing for a while to edit the next character. Press to move leftwards
and to move rightwards. Press to clear.
4. Repeat 2-3 steps to complete the first 16 characters.
5. Press to complete the next 16 characters. The shown is
Arrow “ ” pointing at LO means that the last 16 characters can be edited.
6. Do as steps 2-4
7. Press to save and return to normal weighing mode.
8-11Time Setting
1. Press and hold under normal weighing mode.
2. Press to enter adjusting mode.
3. Press to move leftwards and to move rightwards. Use numeric key to input the
adjustment.
4. Press to save and return.
8-12 Error message
Error Message Reasons Solutions
The CPU unable to read the
EEPROM
Return to the Manufacturer
Unable to read the 3 points
calibration range
Do the 3 points calibration
Zero Point is too high Make sure the pan is empty when turn
on the scale or perform the 3 points
calibration.