Vetek Weighing AB
JCL 9
2) Here we take JCL-15K as an example.
6-1 Single Point Calibration
1) Press and hold key SMPL while powering on
the scale.
2) Input 11 with numeric keys
3) Again press key SMPL to enter the zero point
calibration modes.
4) Wait till flashes, press numeric key 1 to select calibration weight .Options are 1/3
of full load, 2/3 of full load and full load. E.g., options for JCL-15K are 5, 10 and 15(kg).Put the
corresponding weights on the weighing pan.
5) Press key · to confirm.
6) The calibration procedure is completed with a ― flashing on the weight screen. Now,
remove all the weights.
7) Press key SMPL to save, and then press key ZERO to return to the weighing mode.
6-2 Linear Calibration
1) Press and hold key TARE while powering on
the scale.
2) Again press key TARE to enter zero point
calibration mode, with ― ‖ flashing on the
weight screen.
3) Wait till ― appears and flashes on weight
window, put weights of 1/3 of full load on (e.g. 15k
model, 1/3 of full load is 5kg.)
4) Wait till ― appears and flashes on total weight screen, put weights of 2/3 of full load
on (for 15k model, 2/3 of full load is 10kg)
5) Wait till appears and flashes on total weight screen, put weights of full load on (for
15k model, full load is 15kg)
6) The calibration Procedure is finished with a symbol of flashing, and then take
away the weights.
7) Press key TARE to return to weighing mode.