EasyManua.ls Logo

Jadever JCL - 9 Troubleshooting and Error Message

Jadever JCL
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vetek Weighing AB
JCL 13
Data Format
Kg
G/N
.
W
.
+/-
k
g
CR
LF
weight
U
.
W
.
g
/
p
c
s
CR
LF
Unit weight
T
o
t
a
l
p
c
s
CR
LF
pcs
Example
G.W. + 2.2352 kg
U.W. + 0.5352 g/pcs
Total 4176 pcs
lb
G/N
.
W
.
+/-
l
b
CR
LF
weight
U
.
W
.
l
b
/
p
c
s
CR
LF
Unit weight
T
o
t
a
l
p
c
s
CR
LF
pcs
Example
G.W. + 2.2352 lb
U.W. + 0.5352 lb/pcs
Total 4pcs
G = GROSS N = NET
9. Troubleshooting and Error Message
Error message
Problems
Solutions
Initial zero point exceeds
/-30% (take 30% as reference
basis)
1. To check whether there are other alien
articles on the scale pan, remove those
articles.
2. LOAD CELL failure, which requires to be
changed or to contact our Service.
Higher or lower than A/D
resolution range.
1. Check whether it is A/D failure, if yes, please
replace AD.
2. LOAD CELL failure, replacement is required
or contact our Service.