JADEVER JCE/JWE Service Manual - V05
As per the right of proprietary intellectual property, may not be reproduced in any form without written permission.
9
JCE series
u three-point calibration
Follow the procedure below to conduct the three-point calibration.
1. To enter the calibration mode is the first step. (Within the calibration mode, press
to return to the initial step when needed.)
2. Switch off the scale.
3. Press and switch on the scale together, then SET will be displayed.
4. Leave weighing pan empty, and then press again. On0 will be displayed.
JCE starts zero calibration.
5. Then On1 will be displayed after calibration.
6. Put the first calibration weight of 1/3 full capacity onto the weighing pan.
7. Then On2 will be displayed.
8. Put the second calibration weight of 2/3 full capacities onto the weighing pan.
9. Then On3 will be displayed.
10. Put the third calibration weight of full capacity onto the weighing pan.
11. displays and scale beeps.
On0
Set 0 0
On1
On2
On3