JADEVER JCE/JWE Service Manual - V05
As per the right of proprietary intellectual property, may not be reproduced in any form without written permission.
10
12. Calibration is then completed.
13. To return to the normal weighing mode, press .
u span calibration
1. Turn off the scale and leave the pan empty.
2. Keep SMPL key pressed and turn on the scale, 0 will be displayed in UNIT WEIGHT and
TOTAL COUNT displays.
3. Key in the number 11, then press SMPL, to start zero calibration.
CAL 0 -H9-
4. When the symbol CAL blinks, zero calibration is completed and scale is now waiting for
span calibration.
CAL 0 -H9-
5. Press key 1 to chose the weight desired for calibration (1/3, 2/3, or full capacity ).
6. Put on the weight desired for calibration and press , then span calibration will be
executed.
7. If span calibration is completed, the display will show PASS. Remove the calibration
weight from the pan, then press SMPL back to the setting mode.