JADEVER JCE/JWE Service Manual - V05
As per the right of proprietary intellectual property, may not be reproduced in any form without written permission.
16
JWE series
Follow the steps below to enter the parameters setup mode.
1. Turn off the scale.
2. Keep MC/SET pressed and turn on the scale. Display will show CAL
3. Now, you can press MC/SET to enter the power-on setup mode.
4. Press MC/SET again to go to the next setup when current setup is completed.
5. Press ZERO at any stage to go to normal weighing mode
Ø Capacity-resolution setup
1. Press MC/SET right after display shows CAL. Display will show the current
capacity-resolution setup.
2. Press M+/MR repeatedly until the desired capacity-resolution is displayed. (The
available capacity-resolution is model dependent.)
Ø Operating units selection
This function is provided for you to enable those weighing units you want to have in normal
operation. Those disabled units will not be shown when you press MODE in normal
operation.
1. Press MC/SET until display shows on kg.
2. Press M+/MR to view the status of the available units. They are:
Kg / g / lb / lb-oz / (TAEL, China) / 港斤 (TAEL, Hong Kong) / 台斤 (TAEL,
Taiwan) / pcs (piece counting) / % (percentage)
3. Press MODE to enable (ON) or to disable (OFF) each of it.
Ø Power-on operation unit
You can choose one of the weighing units mentioned in the previous section as the power-on
weighing unit. The selected unit will be the default unit each time when the scale is powered
on (other enabled units can be displayed in revolving order by pressing MODE.)
1. Press MC/SET until display shows Init = g.
2. Press M+/MR to choose the unit you want to set as power-on operation unit.
Ø Digital filtering
This function is provided for weighing vibrating objects, or to prevent the scale from vibration
that may affect stability of the scale.