10 USER MANUAL
INSPECT: BULKHEAD (Single Fiber Connectors)
INSPECT
1. Install the appropriate inspection tip and
barrel assembly (FBPP-BAP1)
to the probe microscope.
Note: Standard tips require barrel assembly.
2. Insert the probe into the bulkhead to
inspect.
3. Turn the focus control on the probe to focus
the ber image on the display.
4. Determine whether clean or dirty.
If clean, do not touch it and CONNECT.
If dirty, and if cleaning is required,
CLEAN.
Focus Control
2
3
INSTALL TIP AND BARREL ASSEMBLY
INSPECT BULKHEAD
4
Inspection Tip Barrel Assembly
1