5.6. Adjusting the pump flow
1
2
The pump flow is pre-adjusted in
the factory and can be modified to
suit user requirements.
Using a screwdriver turn the screw (1)
anti-clockwise by 360 °.
Using flat pliers turn the marking of the
slide (2) to the desired position.
Tighten the screw.
Examples:
Internal applications in the bath
A 100 % internal bath circulation
(for large bath tanks)
B Reduced internal bath circulation
(for smooth surface of bath liquid)
External/internal applications
C 40 % external discharge,
60 % internal circulation
(for large bath tanks)
D 80 % external discharge,
20 % internal circulation
(for small bath tanks)