15. Technical specifications
MV MW-Z MW
-4/-6/-12/-26 -4/-6/-12/-26
Working temperature range °C 40 ... 200 25 ... 200 45 ... 200
with tap water cooling °C 20 ... 200 20 ... 200 20 ... 200
Temperature stability ±0.01 °C
Temperature selection digital
via keypad indication on MULTI-DISPLAY (LED)
remote control via personal computer indication on monitor
Temperature indication MULTI-DISPLAY (LED)
Resolution 0.1 °C
ATC - Absolute Temperature Calibration ±3 °C
Temperature control PID
Working temperature sensor Pt 1000
Safety temperature sensor Pt 1000
Heater wattage 2000 W (at 230 V)
or 1000 W (at 115V)
Pressure pump:
pressure, max. at 0 liter 340 mbar 340 mbar 340 mbar
discharge, max. at 0 bar 15l/min 20 l/min 20 l/min
Suction pump:
suction, max. at 0 liter 220 mbar 220 mbar
discharge, max. at 0 bar 14 l/min 14 l/min
Electrical connections
Alarm output 24-0 V DC / max. 25 mA
Computer interface RS232C
Ambient temperature 5 °C ... 40 °C
Mains power connection ±10 % 230 V / 50 Hz
or 115 V / 60 Hz
Total power consumption
(at 230 V) W 2100 2150 2150
(at 115 V) W 1100 1150 1150
All measurements have been carried out at:
rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.