Questions and Answers

  • A
    Ashlee AndersonAug 1, 2025
    What to do if Julabo MV-4 heater and circulation pump shut down?
    • J
      John FordAug 1, 2025
      If the heater and circulation pump of your Julabo Thermostat completely shut down, it could be due to several reasons. First, ensure that the bath tank has sufficient bath liquid. A low liquid level, possibly from a tube breakage causing excessive bath liquid pumped out, can cause this issue; in this case, replace the tubing and replenish the bath liquid. Also, check if the float is defective. The safety temperature value might be set too low; set it to a higher value. A sudden temperature increase, perhaps from an exothermic reaction or preheated samples, can also trigger a shutdown; in this case, set the safety temperature to a higher value. After addressing these potential issues, reset the system by switching the mains power off and on again.
  • J
    Joshua RileyAug 16, 2025
    What to do if Julabo MV-4 shows pump motor overload protection?
    • J
      jaimeshieldsAug 16, 2025
      The pump motor of your Julabo Thermostat is protected against overloading. After a short cooling interval, the motor will automatically start running.

Summary

Safety Warnings

General Safety Precautions

Essential safety instructions and warnings for safe operation and handling of the circulator.

Operating Controls and Functional Elements

Device Components Overview

Detailed identification of physical controls, indicators, and connectors on the circulator unit.

Product Description

Preparations

Operating Procedures

Power Connection and Voltage Check

Requirements for connecting the unit to a power source and verifying voltage.

Switching On, Starting, and Stopping

Steps for powering on, starting, and stopping the circulator operation.

Setting Operating Temperatures

Guide to setting working temperatures T1 and T2 on the circulator.

Setting Warning Limits

Instructions for setting high and low temperature warning limits.

Safety Temperature and Liquid Level Protection

How to set the safety temperature and understand low liquid level protection.

Troubleshooting and Error Messages

Error Code Explanations

Common error codes (E 01, E 05, E 06) and their causes and remedies.

Safety Recommendations

Operational Safety Guidelines

Comprehensive safety advice for operating the circulator and handling liquids.

Absolute Temperature Calibration (ATC)

Electrical Connections

Remote Control Operations

Error Message Directory

Comprehensive list of error messages with their descriptions.

Cleaning and Maintenance

Technical Specifications

Julabo MV-4 Specifications

General IconGeneral
Temperature Stability±0.01 °C
Pump Capacity15 l/min
Pump Pressure0.35 bar
InterfaceRS232
RefrigerantR134a
Pump TypeMagnetic drive
Power Supply230 V