Remote control
26
12.3. List of commands
Command Parameter Response of circulator
version none Number of software version (V X.xx)
status none Status message, error message (see below)
out_mode_01 0
Use working temperature "T1" *
out_mode_01 1
Use working temperature "T2" *
out_mode_05 0 Stop the circulator = r OFF
out_mode_05 1 Start the circulator
out_sp_00 xxx.x Set working temperature "T1"
out_sp_01 xxx.x Set working temperature "T2"
out_sp_02 xxx.x
Set high temperature warning limit
out_sp_03 xxx.x
Set low temperature warning limit
.
in_sp_00 none Ask for working temperature "T1"
in_sp_01 none Ask for working temperature "T2"
in_sp_02 none
Ask for high temperature warning limit
.
in_sp_03 none
Ask for low temperature warning limit
.
in_pv_00 none Ask for actual bath temperature
in_pv_01 none Ask for the heater wattage being used
12.4. Status messages
Message Description
00 MANUAL STOP Circulator in "OFF" state
01 MANUAL START Circulator in keypad control mode
02 REMOTE STOP Circulator in "r OFF" state
03 REMOTE START Circulator in remote control mode
(* see "Note" on page 21).