Troubleshooting guide / Error messages
18
7. Troubleshooting guide / Error messages
!
+
Whenever the microprocessor electronics registers a failure, a
complete shutdown of the heater and circulation pump is
performed. The alarm light " " illuminates and a continuous
signal tone sounds.
Cause Remedy
The circulator is operated
without bath liquid, or the liquid
level is insufficient.
Replenish the bath
tank with the bath
liquid.
Tube breakage has occured
(insufficient filling level due to
excessive bath liquid pumped
out).
Replace the tubing and
replenish the bath tank
with the bath liquid.
The float is defect (e. g., because
damaged in transit).
Repair by authorized
JULABO service
personnel.
The safety temperature value lies
below the working temperature
setpoint.
Set the safety
temperature to a higher
value.
A heat reaction or sudden
temperature increase, e.g. caused
by an exothermic chemical
reaction or by the immersion of
preheated samples.
Set the safety
temperature to a higher
value.
After eliminating the malfunction, press the mains power switch
off and on again to cancel the alarm state.
The wires of the working temperature sensor are interrupted
or short-circuited.
Defect of the working or safety temperature sensor.
The difference between the working temperature and safety
sensors is above 25 °C.