K-Commander Loss-in-Weight Programming
Rev: G Produced by the K-Tron Institute
LWF Page 04: Alarm Limits, Continued
Page 04
variable list
con't
Variable Name Edit? Description
Clear Alarms on Stops ` This selection determines whether
the stop input or run enable input
will clear alarms or not. If enabled,
alarms will automatically be
cleared. If disabled, the alarms will
be cleared by the user through the
keyboard or by the Alarm
Shutdown Input.
The menu is below:
Disabled
Enabled <
Net Weight Low Limit `
This entry sets the level where the
low weight limit alarm is detected.
The value of weight at which this
alarm is presented is;
Weight = % ( Scale Range - Tare)
ex: SR = 120 kg, Tare = 50 Kg, % = 4
Weight = 4%(120-50) = 2.8 kg
Range of Values
0-10%
Default value: 2%
On Rev. Z software, this value will
also set the under limit for Scale
Under-range using this value a -%
value so that scale under-range
occurs at this percentage below 0
gross weight.